Advance multiple microcontroller based generator for Open and Laparoscopic Surgeries
5-inch LCD touch screen display with intuitive interface, and additional feather touch buttons
Intuitive User Interface facilitates easy program and modes selection with Seamless Operations
Features 2 Monopolar, 1 Bipolar and I Ligate. Socket, for handswitch and footswitch activation
It is compatible with universal accessories used in Monopolar and Bipolar Surgical Applications
Supports ALAN’s special Capacitive Patient Plate for maximum safety of the patients in surgeries
PPCM feature monitors patient’s contact quality and provides real-time graphical representation
“Self-Test”, diagnoses at every “START UP” for Generator’s Healthiness & Operational Integrity
Automatically stops the power output and provides audio-visual indication during problem
Multiple cutting modes with varied regulated wattages for smooth and precise skin incisions
Instant Plasma Generation feature allows the surgeon to vaporize tissues giving cutting effect
Larger Prostate Glands can be resected without TUR Syndrome in Saline based TURP Surgeries
Simultaneous monopolar coagulation for 2 surgeons via handswitch and footswitch pairing
Auto feature in Monopolar & Bipolar modes ensure optimum power, without tissue charring
System restores previous activated power settings when encountered power interruptions
HF power output can be adjusted using touch screen and handswitch pencil from sterile field
“Current Meter Display” in the Bipolar Mode to Indicate the “Flow of Current” during surgeries
HF current leakage monitoring & controlling system supervises and controls it less than 10 uA
99+1 user programmable memory for surgeons. to store & recall surgery preferences and details
Upgradeable and compatible with Argon Plasma Coagulation & Smoke Evacuator System
Description
Monopolar Cut Modes:
- Pure : 400 W @ 350 Ohms CF 1.5
- Lapcut : 300 W@350 Ohms CF 1.5 Input
- Blend : 250 W @ 350 Ohms (10 Levels) CF Variable
- EndoCut : 300 W @ 350 Ohms (10 Levels) CF Varied
Monopolar Coagulation Modes:
- Spray : 120 W@350 Ohms CF 9.0
- Fulgurate : 150 W @ 350 Ohms CF 7.5
- Desiccate : 200 W @ 350 Ohms CF 3.0
- Soft : 300 W @ 150 Ohms CF 1.5
- Auto Desiccate : Monopolar Desiccate with Auto-Stop
- Auto Soft : Monopolar Soft with Auto-Stop
Bipolar Cut Modes:
- LoCut : 120 W @ 100 Ohms CF 1.5
- HiCut : 120 W @ 100 Ohms CF 1.5
- ScalCut : 200 W @ 100 Ohms (5 Levels) CF 1.5
Bipolar Coagulation Modes:
- Cut/Standard : 120 W @ 100 Ohms CF 1.5
- Macro/Force : 120 W @ 100 Ohms CF 1.5
- Micro/Precise : 100 W @ 100 Ohms CF 1.5
- Auto Micro : Bipolar Micro with Auto-Stop
- Auto Start : Auto-Start, Auto-Stop
Bipolar Saline Resection:
- LowCut : 250 W @ 200 Ohms CF 1.5
- HiCut 1 : 300 W @ 200 Ohms CF 1.5
- HiCut 2 : 300 W @ 200 Ohms CF 1.5
- Effect 1 : 300 W @ 200 Ohms CF 1.5
Bipolar Saline Coagulation:
- Desi : 250 W @ 25 Ohms CF 1.5
- Force : 250 W @ 25 Ohms CF 2.1
Input Supply
- Input : 88-265V AC, 50/60 Hz
- Power Consumption : 750VA Max
- Input Fuse : 5A T
- Output Control : Colour TFT Display with Touch Screen and Push buttons in all Modes
- Max. Power Output : 400 Watts
Safety Standards
- Basic Construction : As per IEC 60601-1, IEC 60601-1-2 & IEC 60601-2-2
Classifications
- Protection Class : I
- Mdd. Class : II b
- Pollution degree Class : II
- Overvoltage category Class : II
- Unit Type : CF
- All Outputs : Floating Output (All output ports are CF Type)
- Electrical Potential Balancing : Connecting Pin and Indicated by Symbol
- PER : 98
- Dimension : 335mm (W) x 160mm (H) x 395mm (D)
- Weight : Maximum 8.0 Kg
Operating Parameters
- Ambient temperature range : +10° C to +40° C
- Relative humidity : 30% to 75% non-condensing
- Atmospheric pressure : 700 millibars to 1060 millibars
Transport and Storage
- Ambient temperature range : -30° C to +65° C
- Relative humidity : 0% to 90% non-condensing
- Atmospheric pressure : 500 millibars to 1060 millibars